EMC Compliance
Table C-2: EMC Immunity Guidance and Declaration
IEC 60601-1-2
Immunity Test
Test Level
Electrostatic discharge
Contact: ± 6 kV
(ESD)
Air: ± 8 kV
IEC 61000-4-2
Electrical fast
Power supply
transient/burst
lines:
IEC 61000-4-4
± 2 kV
Input/output
lines:
± 1 kV
Surge
Line-line: ± 1 kV
IEC 61000-4-5
Line-earth: ± 2 kV
Voltage dips, short
< 5% U
interruptions and
(> 95% dip in U
voltage variations on
for 5 sec
power supply input
lines
IEC 61000-4-11
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Compliance
Level
Contact: ± 6 kV
Air: ± 8 kV
Power supply
lines:
± 2 kV
Input/output
lines:
± 1 kV
Line-line: ± 1 kV
Line-earth: ± 2 kV
< 5% U
T
T
)
(> 95% dip in U
T
for 5 sec
*
Electromagnetic Environment Guidance
Floors must be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity must be
at least 30%.
Mains power quality should be that of typical commercial or
hospital environment.
Mains power quality should be that of typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the system requires
)
T
continued operation during power mains interruptions, it is
recommended that the system be powered from an
uninterruptible power supply or a battery.
C-2