D670/Opt Reduced Service Clearance Configuration - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

Equipment Description and General Construction Requirements

2.2.2.1 D670/OPT Reduced Service Clearance Configuration

NOTE
D670/OPT
This section is relevant only for D670/OPT systems.
In the reduced service clearance configuration, service, egress and workspace around the
gantry are compromised.
No permanent objects that reduce the CT service area around the gantry are allowed in this
reduced minimal configuration.
When installing D670/OPT in the reduced room configuration, the small-room 14"-28" option
(B7700SB) should be ordered.
IMPORTANT
The reduced service clearance configuration puts severe limitations on space for patient care and work flow.
Customer approval of site drawings signifies customer agreement to these limitations.
CAUTION
!
In the reduced service clearance configuration, the customer should consider workflow, customer access for
patient care, and critical-care operations space requirements. Additionally, this configuration may offer only
limited equipment access on the CT gantry left side when loading patients or when positioning patient equipment
in the room between the gantry and the wall.
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare