Equipment Description and General Construction Requirements

2.2.3 System Mechanical Curves

Figure 2-14: D670 Component Movement Curves, p.2-22
movement. In addition, the contrast injector, ECG trigger monitor and collimator carts can be
moved to different locations in the scan room, as demonstrated in the layout illustrations in
Figure 2-10, p.2-17
NOTE
In order to prevent collision with the gantry display boom, do not mount any equipment from the ceiling.
Table slanted at
°
67.5
°
55
°
42.5
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
through
Figure 2-13,
p.2-20.
Farthest point relative to system's center line
190.5 cm (75")
181.5 cm (71.5")
157.5 cm (62")
2-21
2.2 Room Size, Layout and Considerations
2.2.3 System Mechanical Curves
illustrates the table and gantry