Equipment Description and General Construction Requirements
Table 2-4: Floor Leveling Specifications
Item
Floor leveling area
Slope
Floor surface
Flatness
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Requirement
640 cm×370 cm (21'×12') (covering the entire planned area of table and gantry installation,
depending on room layout)
3 mm (0.125") over 3048 mm (120")
A single poured surface
The surface must be smooth and without significant valleys or peaks.
The entire surface area must have an overall flatness of 5 mm over 1500 mm in any direction
(see
App.B, Measuring Floor Flatness
5.1
5.1
2
2
5
4.8
t
for measurement procedure).
5.3
4.9
2.1
1.8
5.2
5
2-40
2.3 Room Structural Requirements
2.3.1 Floor Requirements
5.5
5.3
Slope
= Pass
Flat = Fail
2.4
2.1
Slope
= Fail
Flat = Fail
5.3
5.2
Slope
= Fail
Flat = Pass