Equipment Description and General Construction Requirements
Table 2-2: Weight of Components(Continued)
Component
LEHR collimator
LEGP collimator
ELEGP collimator
MEGP collimator
HEGP collimator
NM UPS
CT UPS (optional)
Xeleris workstation
If the system is installed on a floor type thinner than a 166 mm (6.5") concrete floor, the customer shall,
at their expense, provide acceptable anchoring and mounting methods that meet all structural specifications
provided in sections
The customer shall ensure that the floor strength in the collimator cart storage area and along the movement
routes for collimator exchange are suitable for the collimator cart load (approx. 250 kg each).
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Weight (kg)
Weight
(lb)
62
55
62
103
131
May vary but no
May vary but no
more than 130
more than 287
350
(insignificant)
CAUTION
!
2.3.1.2
and
Load Distribution
137
121
137
227
288
4 feet
772
4 wheels; Anchored to floor
2.3.1.3
of this manual.
2-29
2.3 Room Structural Requirements
2.3.1 Floor Requirements
Comments
2 per system/cart
2 per system/cart
2 per system/cart
2 per system/cart
2 per system/cart