D670 Required Corridor Width For 90° Turns - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

General System Requirements
Table 1-1: D670 Components and Clearance — Metric(Continued)
Component name
Optional Items
NM UPS
ECG Trigger Monitor
Xeleris
Monitors
NM600 Detectors Dismount Option
NM gantry without detectors
Detector 1
Detector 2
*
The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor leading
to the room. When planning or measuring the width of the scan room door, use the graphs provided in
for Corridor and Scan Room Door, p.1-17
The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See
May be delivered a few days prior to system delivery, as part of the final room check and preparation for installation.
** 20 mm clearance above the floor
†† Weight of gantry in the in-site transportation configuration: 1330 kg + weight of the dolly: 310 kg
‡‡ 50 mm clearance above the floor
*** The specified weight includes the packing. The detectors must be conveyed crated (unpacking is allowed only at the room or designated area
where the detectors to be installed)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Crated
Crate size
Weight
(Dollies not attached)
(cm) (Height × Width ×
Depth/Length)
May vary but not
May vary but
more than
not more than
60×40×80
150
May vary but not
May vary but
more than
not more than
80×80×80
15
220×150×168
2125
93×86×100
320
93×86×100
320
in order to verify that the measurements comply with the requirements.
Figure 1-3: D670 Required Corridor Width for 90° Turns, p.1-18
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Minimal dimensions (cm)
Corridor /
Corridor /
(kg)
Door
elevator
elevator
width
width
length
any
any
any
any
any
any
94.5
94.5
222
86
86
100
86
86
100
1-11
1.3 Delivery Requirements
Uncrated
*
Weight
Width of
corridors with
Height
90° turns
May vary but
any
any
any
any
**
188
195
‡‡
100
98
‡‡
100
98
Figure 1-2: D670 Relative Required Width
for the required width when the angle is 90°.
(kg)
not more
than 130
<13
††
1640
(with dollies)
***
320
***
320

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents