Magnetic Field Considerations - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

Special Construction Requirements

3.2 Magnetic Field Considerations

The ambient static magnetic field in the system location must be less than 10
milligauss). The ambient AC magnetic fields must be below the 10
IMPORTANT
The system must be installed in an x-ray protected room providing an attenuation of at least 12 db for radio disturbances
from 30 MHz to 1 GHz.
Low Frequency Magnetic Field
N/A
Static Magnetic Field Limits
In order to avoid interference on the system, the static field limits from the surrounding environment
must be less than 1 Gauss in both the scan and the operator rooms.
s
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
3-6
3.2 Magnetic Field Considerations
-4
tesla (1,000
-6
tesla (10 milligauss) peak.