Chapter 6: Network Requirements - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

Chapter 6: Network Requirements

The system requires the following network connections:
Broad-Band Network Connection (BBNC) (required): broad-band network connection wall jack,
located within 1 m (39") of console location, for internal hospital networking and InSite
broadband connectivity.
Local Area Network (LAN) (required)
LAN connections are usually required in the operator room for:
The LAN and WAN Networks sockets/outlets (minimum 3) must be available in the
operator room within a distance of 1 m (39") from the designated location of the operator
console, processing workstations (Xeleris or A/W) and LAN printer installed in the
operator room.
In the scan room it is recommended to have one LAN socket/outlet available in close
proximity to the CT or NM gantry for service engineer activities actions.
Wide Area Network (WAN) (optional)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Xeleris or A/W processing station
CT subsystem
NM subsystem
DICOM LAN printer (optional)