Equipment Description and General Construction Requirements

2.3.5 Vibration Specifications

The system components are sensitive to vibration in the frequency range of 0.5 to 20 Hz, depending
on the amplitude of the vibration. It is the customer's responsibility to contract a vibration consultant
or qualified engineer to verify that these specifications are met and implement an appropriate
solution.
To minimize vibrations, the system must be installed on a solid floor, as far as possible from the
following vibration sources:
Outside building
Parking lots
Roadways
Subways
Heliports
Trains
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Inside building
Hallways
Elevators
2-44
2.3 Room Structural Requirements
2.3.5 Vibration Specifications
Other
Hospital power plants
containing pumps, motors, air
handling equipment and air
conditioning units