Environmental HVAC Requirements
Table 4-1: Requirements for Ambient Temperature, Humidity and Altitude
Temperature
Humidity
Altitude
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Maximum
Minimum
26°C (79°F)
18°C (64°F)
60%
30%
non-condensing
non-condensing
relative humidity
relative humidity
2400 m (7,875 ft.)
-150 m (-492 ft.)
Recommended
22°C (72°F)
4-4
4.1 General Guidelines
Maximum rate of change
3°C/hr (5°F/hr)
5%/hr