Equipment Description and General Construction Requirements
Defines the recommended
room dimensions required to
enable installation, operation
and service of the system in
safe conditions.
Does not take into account
local requirements.
Enables relative freedom of
movement for operator and
service.
Enables installation of table
extender.
Can keep two collimator carts
in the room.
The operator room is not taken
into account in the room
dimensions.
Space must be allocated for
the operator room, but the
location can differ.
Note:
Dimensions are in mm
Figure 2-11: Recommended Layout
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Pivot – 2000
6450 - Footprint
250
2-18
2.2 Room Size, Layout and Considerations
2.2.2 System Layout Drawings
Optimal + Extender Layout
50+100 400 FOV
1150 – NMFOV to CT
6750
50
350
1600 – CT Scan Length