Equipment Description and General Construction Requirements

2.2 Room Size, Layout and Considerations

The system requires a main Scan Room and a separate adjoining Operator Room, which contain
the following sub-systems:
Table 2-1: D670 Components in Scan and Console Rooms
Scan Room
Fixed Components
(see
Figure 2-1,
NM gantry
Patient table
CT gantry
UPS (optional)
CT PDU
MDP
PIDB
EMO
This section provides guidelines for determining the size and layout of the scan and operator rooms
and of the above components, and example layouts of typical rooms, illustrating the position and
dimensions of the components.
The room layouts provided take into consideration all aspects of operation, operator and patient
requirements and service clearance requirements.
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Moving Components
p.2-3)
(see
Figure 2-2,
p.2-4)
Collimator carts
ECG trigger monitor
Contrast injector
Operator Room
(see
Figure 2-3,
p.2-5)
NM acquisition station
Xeleris workstation (optional)
FWS table (optional)
Operator chair
CT console
AW workstation (optional)
2-11
2.2 Room Size, Layout and Considerations
Variable location
Storage cabinet
(not supplied with the
system)