Equipment Description and General Construction Requirements
Safety zone considerations:
The safety zone is designated by tape on the floor, usually
yellow (can also be differentiated by a change in floor
coloring). This designates the area that must be free of
obstructions to avoid a collision during an automatic motion.
Operation-related considerations:
Enable access for hospital beds, including maneuvering
and positioning the bed and moving the front of the
patient table during collimator exchange.
Storage of the collimator carts (one for each set of
collimators) when not in use
ECG Trigger Monitor and Contrast Injector – cable
position and lengths and storage when not in use
Space for storage and usage of ECG Trigger Monitor and Contrast Injector
Installation and service considerations:
Location of power connections
Access to communication lines (Ethernet, external hardcopy device)
Floor loading capacity and weight of system components, including storage and path of collimator carts
Service clearance areas (see
Storage cabinet for storage of service tools (optional). Depending on the room layout, it is
recommended that sufficient area is allocated for the cabinet.
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
App.D, Regulatory
Clearances)
2-24
2.2 Room Size, Layout and Considerations
2.2.4 Layout Considerations
Detector
Pallet
motion area
motion area