Equipment Description and General Construction Requirements
Table 2-5: Subsystem Centers of Gravity and Anchoring Points
IMPORTANT
For special seismic kit details and information, refer to the specific ITF released for the system seismic install.
Component
Patient table
NM gantry
CT gantry on floor plates
CT gantry in service position
NIO Console
UPS
CT PDU
NM acquisition station
Collimator carts
*
In the most unfavorable configuration
† http://powerquality.eaton.com/Products-services/Backup-Power-UPS/9155-info.asp
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Center of Gravity Location
(cm)
X
Y
Z
109
30
50
19.1
102*
0
*
*
*)
(+107.5
109.5
*
*
110
23.8
30.6
10
35
23.5
49.5
48.5
26.5
99
Anchoring Method
Anchor plate + 6×HITLI-HLC 6.5×60/40
4 × HILTI HSL-3-G M10/100 anchors
7× HILTI HSL-3 M10/40 anchors
None
None
None
None
None
Carts cannot be anchored, as they must move
freely in the room for collimator exchange.
2-47
2.4 Seismic Requirements
See also
Figure 2-16, p.2-31Figure
2-21, p.2-38
Figure 2-20, p.2-36
Figure 2-20, p.2-36
Figure 2-17, p.2-32
Figure 2-6, p.2-8
Figure 2-7, p.2-9