General Transportation And Delivery Precautions; General Temperature Precautions - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

General System Requirements

1.3.1 General Transportation and Delivery Precautions

1.3.1.1 General Temperature Precautions

Extreme temperatures must be avoided during system transportation and delivery. Ensure that
the system is not exposed, for an extended period of time, to temperatures or humidity outside
the following specifications.
Temperature range: -34ºC to +60ºC (-29ºF/+140ºF)
Humidity range (non-condensing): 5% to 95%
NOTE
Component freezing occurs if the system is exposed to temperatures below -18
than two days. Allow a minimum of 12 hours for the system to adjust to ambient room temperature, prior to
installation.
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1.3.1 General Transportation and Delivery Precautions
1-5
1.3 Delivery Requirements
º
º
C (0
F) for a period of longer

Hide quick links: