General System Requirements
Table 1-3: D670 Components and Clearance — Imperial(Continued)
Component name
ECG Trigger Monitor
Xeleris
Monitors
NM600 Detectors Dismount Option
NM gantry without
detectors
Detector 1
Detector 2
*
The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor
leading to the room. When planning or measuring the width of the scan room door, use the graphs provided in
Required Width for Corridor and Scan Room Door, p.1-17
The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See
May be delivered a few days prior to system delivery, as part of the final room check and preparation for installation.
** 20 mm clearance above the floor
†† Weight of gantry in the in-site transportation configuration: 1330 kg + weight of the dolly: 310 kg
‡‡ 50 mm clearance above the floor
*** The specified weight includes the packing. The detectors must be conveyed crated (unpacking is allowed only at the room or designated area
where the detectors to be installed)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Crated
Crate size
"
(
)
Weight
(Dollies not attached)
(lb)
(Height × Width ×
Depth/Length)
May vary but not
May vary
but not
more than
more
31.5x31.5x31.5
than 33
86.6×59×66.1
4685
36.6×33.8×39.4
705
36.6×33.8×39.4
705
in order to verify that the measurements comply with the requirements.
Figure 1-3: D670 Required Corridor Width for 90° Turns, p.1-18
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Uncrated
Minimal dimensions (")
Corridor /
Corridor /
Door
elevator
elevator
width
width
length
any
any
any
37.2
37.2
87.4
33.8
33.8
39.3
33.8
33.8
39.3
for the required width when the angle is 90
1-14
1.3 Delivery Requirements
*
Weight
Width of
corridors with
Height
°
90
turns
any
any
<28.6
3616
**
74
76.7
(with dollies)
‡‡
39.3
705
38.5
‡‡
39.3
38.5
705
Figure 1-2: D670 Relative
(lb)
††
***
***
°
.