Customer Checklist
Table A-3: Room Preparation (Continued)
Description
Environmental
conditions
Room shielding
Safety
Customer-Provided
Cabling and Connections
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Ample working light is available for service
Air-conditioning meets requirements for system
thermal loads
Air-conditioning meets humidity requirements
Magnetic field in camera room is < 1 Gauss
Room is clean and free of dust, ready for installation
Shielding of scan room meets requirements
Shielding of operator room meets requirements
External exposure x-ray light installed
Planned location of emergency button in scan room
is easily accessible by operator
Interlock system installed
Door switch
Warning lights
System Emergency Off button (SEO)
Additional wiring from MDP to PDU and UPS
available
Status See
A-5
Comments