Equipment Description and General Construction Requirements
Table 2-3: D670 Drilling and Anchor Chart(Continued)
No.
X
9
-980.00
10
-980.00
11
-1495.00
12
-1495.00
13
-1880.00
14
-1495.00
15
-980.00
16
-980.00
17
-1495.00
18
-1965.00
19
-2055.00
20
-2545.00
21
-2065.00
22
-2545.00
23
-2545.00
24
-2065.00
25
-2065.00
26
-2545.0
27
-1703.50
28
-1623.50
29
-1623.50
30
-1703.50
31
-349.00
32
-279.00
33
-279.00
34
-349.00
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Drill
Hole
Anchored
Y
Hole
Depth
Part
835.65
630.65
CT Short
630.65
855.65
-630.65
-630.65
-830.65
CT Long
-855.65
-855.65
-855.65
966.85
891.85
CT Service
Plate -1
891.85
102.0
966.85
-891.85
-966.85
CT Service
Plate -2
-891.85
-966.85
72.50
Bearing Plate
-72.50
of Service
72.50
Lead Screw
-72.50
63.50
Gear Bracket
-63.50
of Service
63.50
Lead Screw
-63.50
Drilling
Hole Purpose
Method
Main Anchor
Plate
Alternative Anchor
Main Anchor
Plate
Alternative Anchor
Main Anchor
Main Anchor
Alternative Anchor
Main Anchor
Alternative Anchor
Main Anchor
During
Alternative Anchor
Main Anchor
Installation
Alternative Anchor
2-37
2.3 Room Structural Requirements
2.3.1 Floor Requirements
Anchor
Torque
Type
Nm.
HILTI
HSL-3
35
M10/20
Lead
Screw
Section
Figure 2-3