Appendix C: EMC Compliance
This equipment complies with IEC60601-1-2 Edition 2 EMC standard for medical electrical
equipment.
The system is suitable to be used in an electromagnetic environment, in compliance with the limits
and recommendations provided in the following tables:
Emission compliance level and limits
Immunity compliance level and recommendations to maintain equipment clinical utility
Table C-1: EMC Emission Declaration
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-2
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Compliance
Group 1
Class A
NA
NA
C-1
Electromagnetic Environment Guidance
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The system is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
-