General System Requirements
Table 1-2: D670 Additional CT-specific Components and Clearance — Metric
Crate size
Component
(Dollies not attached)
name
(cm) (Height × Width ×
CT PDU
123×90×70
CT UPS
CT console (NIO)
CT accessories
115×100×150
CT sub-system
190×100×170
covers
MDP (Main
100×70×50
‡ **
Disconnect Panel)
Optional Items
Contrast Injector
May vary but not
AW W/S
*
The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor leading to
the room. When planning or measuring the width of the scan room door, use the graphs provided in
Corridor and Scan Room Door, p.1-17
The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See
A1 or other; if ordered from GE, delivered in advance as part of the site preparation and must be installed prior to system installation
** In some cases, an alternative MDP is supplied, in order to comply with local electricity standards, see
(A1), p.5-14
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Crated
Weight
Depth/Length)
D670/OPT 460
D670/BSE 440
82×31×81
65×47×83
115
50
537
60
May vary but not
more than
more than 15
80×80×80
in order to verify that the measurements comply with the requirements.
Figure 1-3: D670 Required Corridor Width for 90° Turns, p.1-18
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Minimal dimensions (cm)
Corridor /
(kg)
Door
elevator
width
width
100
100
any
any
any
any
100
110
80
80
any
any
any
any
1-12
1.3 Delivery Requirements
Uncrated
*
Width of
Corridor /
corridors with
elevator
Height
°
length
90
turns
100
100
any
any
any
any
any
any
any
150
180
any
100
128
199
any
any
any
any
any
any
Figure 1-2: D670 Relative Required Width for
for the required width when the angle is 90°.
Power Input Distribution Box (PIDB) with MDP
Weight
(kg)
D670/OPT 370
D670/BSE 350
160
88
50
30
45
<13