EMC Compliance
Table C-2: EMC Immunity Guidance and Declaration (Continued)
IEC 60601-1-2
Immunity Test
Test Level
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
3
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3
IEC 61000-4-3
80 kHz to 2.5 GHz
(alternative method:
IEC 61000-4-21)
*
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
UT is the AC mains voltage prior to application of the test level.
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Compliance
Level
3 A/m
3
VRMS
VRMS
150 kHz to 80 MHz
3
V/m
V/m
80 kHz to 2.5 GHz
*
Electromagnetic Environment Guidance
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Note:
U
is the AC mains voltage prior to application of the
T
test level.
Portable and mobile RF communications equipment should be
used no closer to any part of the system, including cables,
than the recommended separation distance calculated from
the equation appropriate for the frequency of the transmitter.
For recommended separation distances, see
C-3
Table C-3,
p.C-4.