Table 11 File data-mode stored measurements (continued)
File data mode
MIN/MAX
TIME OF FLT
REDUCTION RT
The default file data mode can be set to the most frequently used.
To change the default file data mode
1.
In the measurement screen, press [SETUP], and then select SYSTEM.
2.
In the SYSTEM screen, set DEFAULT FILE MODE to the desired option (see
Table 11 on page 104 for details).
3.
Press [MEAS] to return to the measurement screen.
6.7.4

Performing File Operations

Pressing [FILE] opens a menu from which numerous file operations can be performed
(see Figure 6-26 on page 106). The following sections describe how to perform the
operations. Datalogger files are stored in the internal microSD memory card, and
these files can be imported/exported from/to an external microSD memory card.
Stored measurements
Minimum thickness
Maximum thickness
Time of flight
Material thickness
Reduction rate
DMTA-10022-01EN, Rev. C, January 2015
When to use
When using the MIN/MAX mode
(see "Using the Minimum,
Maximum, or Min/Max Thickness
Mode" on page 129).
When measuring time of flight
When the reduction rate
differential mode is activated (see
REDUCTION RT in "Using
Alarms" on page 131)
Software Options
105