DMTA-10022-01EN, Rev. C, January 2015
COMM PROTOCOL
OUTPUT FORMAT
B-SCAN OUTPUT
FTP OUTPUT
OUTPUT TYPE
MASTER RESET
Performs the measurement reset and the internal memory reset in a single step.
The master reset permanently deletes all stored thickness readings/waveforms that
are stored in the internal memory card of the 45MG.
To activate a reset function
1.
In the measurement screen, press [SETUP], and then select RESETS.
2.
In the RESETS screen (see Figure 12-1 on page 203):
a) In the RESETS list, highlight the desired reset function.
A warning message appears, indicating the type of data that will be reset.
b) Select RESET.
202
Chapter 12
Table 15 Default communication settings
Parameter
CAUTION
Default value
MULTI CHAR
F1
OFF
45MG
FTP