Table 26 Setup Parameter Description - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015
Name
MEAS TYPE
PROBE TYPE
PULSER
POWER
MAX GAIN
INIT GAIN
TDG SLOPE
MB BLANK
ECHO
WINDOW
ECHO 1
DETECT
ECHO 2
DETECT
IF BLANK
M3 BLANK
VELOCITY
ZERO
214
Appendix A
Table 26 Setup parameter description (continued)
Description
Special measurement modes
Transducer types
Pulser power
Maximum receiver gain
Initial TDG gain
Time Gain slope (default)
Main bang blank
Echo detect gate which begins at
end of MB (main bang) blank in
mode 1, or interface echo in
modes 2 and 3. The value reported
for the end of the echo window is
relative to the main bang.
Detection polarity of first echo
Detection polarity of second echo
Blank after interface echo
Blank after first measured back-
wall echo in mode 3
Ultrasonic sound velocity of
material to be measured
Zero calibration factor
Units/Resolutions/Range
Standard
First Peak (with Single Element option)
Dual element
Direct contact (with Single element
option)
Delay line
Immersion
60 V, 110 V, 150 V, or 200 V
0.0 dB to 99.0 dB, 0.3 dB steps
0 to Max Gain, 1 dB steps.
0.0 dB/s to 39.9 dB/s
0 ns to 225 µs
0 ns to 224.71 µs, 55 ns, or MB blank
time interval, whichever is less.
+ or −
+ or −
0 µs to 20 µs
0 µs to 20 µs
0.508 mm/µs to 18.699 mm/µs
(0.0200 in./µs to 0.7362 in./µs)
0.00 to 999.99