Figure 5-3 Selecting A Default Single Element Transducer Setup - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

4.
For the Single Element software option and a single element transducer, load an
appropriate setup:
a) Press [2nd F], [FREEZE] (XDCR RECALL).
b) In the menu, select the default choice for the probe type that you use (ex.:
DEFAULT SINGLE ELEMENT).
c)
In the DEFAULT screen for the type of probe you are using (see the example
shown in Figure 5-3 on page 53), highlight the setup in the list of available
default setups for the transducer that you are using.
Default setup naming convention:
D E F M 1 - 2 0 . 0 / M 1 1 6

Figure 5-3 Selecting a default single element transducer setup

NOTE
The setups listed as USER-1 through USER-35 can be renamed for special
applications. Refer to "Custom Setups for Single Element Transducers" on page 165
for more details on setups.
DMTA-10022-01EN, Rev. C, January 2015
Probe part number (ex. M116)
Probe frequency (ex. 20.0 MHz)
Mode: 1, 2, 3
-
-
Material: M
Metal, P
Plastic
DEF: Default
Basic Operation
53