12.10 Resolving Measurement Problems
Symptom
No echoes or weak echoes,
and no measurement (LOS)
Strong echoes, no
measurements
Strong echoes, incorrect
measurements
208
Chapter 12
Table 16 Measurement troubleshooting
•
There is insufficient couplant, especially on
rough or curved surfaces.
•
The gain is set too low.
•
The material is too attenuating, or the surfaces
are unparalleled or too rough.
•
The gage requires service; try the master reset.
•
The echo is too low in amplitude to be
detected. Try increasing the gain.
•
The echo may be in a blank area of the
waveform, and cannot be detected.
•
The echo is outside of the echo window.
•
Out of calibration; perform calibration.
•
Differential mode — check DIFF flag.
•
Min or Max mode — see "Using the
Minimum, Maximum, or Min/Max Thickness
Mode" on page 129.
•
The material is too grainy, has flaws,
inclusions or laminations, or very high surface
noise; try manual gain adjustment or
extended blank.
Possible explanation