NOTE
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the range parameter,
and the [] and [] keys to adjust the range. Press [MEAS] to stop adjusting the
range.
6.4.3.2

Adjusting the Delay Value

The delay of the waveform display adjusts the beginning of the horizontal span. The
delay can be adjusted to display the waveform of interest in the center of the
waveform display. This function is very useful when using delay line or immersion
transducers, or when measuring thick material, in order to ensure that the measured
echoes can be seen in greater detail.
To adjust the delay value
 In the measurement screen, press the [] and [] keys to adjust the delay.
NOTE
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the delay parameter,
and the [] and [] keys to adjust the delay. Press [MEAS] to stop adjusting the delay.
6.4.3.3

Activating the Zoom Function (Available Only with the Waveform Option)

The zoom function automatically and dynamically sets the delay and range values to
optimally track and show the detected echo in the waveform display.
To activate the zoom function
1.
In the measurement screen, press [SETUP], and then select DISPLAY.
2.
In the DISPLAY screen, set the ZOOM OPTION to ON.
The zoom flag ( ) appears on the lower right side of the waveform display, below
the range parameter.