DMTA-10022-01EN, Rev. C, January 2015
SINGLE CHAR: Single character command is normally used when an
external program is controlling the instrument by sending remote
commands that mimic keystrokes.
b) In OUTPUT FORMAT, select the format of the data being output (F1, F2,
F3,...F10).
NOTE
Contact Olympus for more information on the following communication parameters:
Multi and single character remote commands.
Send formats (F1, F2, F3, F4, F5, F6, F7, F8, F9, and F10).
c)
Set B-SCAN OUTPUT to ON to output the B-scan data when communicating
with the GageView interface program. Set it to OFF when communicating
with other interface programs that do not support the B-scan data. This
parameter only applies to files that have stored B-scan images.
d) Set FTP OUTPUT to 45MG in order to use the 45MG file protocol.
OR
Set FTP OUTPUT to 38DLP in order to use the 38DL PLUS file protocol.
OR
Set FTP OUTPUT to MG2 to use the MG2 file protocol.
e)
Set OUTPUT TYPE to FTP for communication with the GageView software
using the standard file transfer protocol.
OR
Set OUTPUT TYPE to CSV to output in a generic comma-separated variables
(CSV) format, which can be integrated into custom software.