Creating A Data File - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015
6.7.2

Creating a Data File

The following procedure describes how to create a data file in the 45MG.
NOTE
It is also possible to create a 45MG data file from a computer using the GageView
interface program. Refer to the GageView Interface Program — User's Manual (P/N:
910-259-EN [U8778347]) for details.
To create a data file
1.
In the measurement screen, press [FILE], and then select CREATE.
2.
In the CREATE screen (see Figure 6-17 on page 93):
a) In the FILE NAME parameter, enter the desired file name using up to
32 characters.
b) In the DESCRIPTION parameter, optionally enter a description of the
contents of the file.
c)
In the INSPECTOR ID, optionally enter an identification of the inspector.
d) In the LOCATION NOTE parameter, optionally enter an identification of
where the measurements are performed.
e)
Select the appropriate data FILE TYPE for your application (see "Data File
Types" on page 93 for details).
f)
Select the appropriate FILE DATA MODE for your application (see "File Data
Modes" on page 104 for details).
g) Set the DELETE PROTECTION mode to ON or OFF.
Delete protection locks the file so it cannot be deleted. You can unlock the file
for deletion using the file edit function.
h) Depending on the selected FILE TYPE, refer to the following sections for the
remaining steps of this procedure:
INCREMENTAL
SEQUENTIAL
SEQ+CUSTOM PT
92
Chapter 6
see "Incremental Data File Type"
on page 94
see "Sequential Data File Type" on
page 96
see "Sequential with Custom Point
Data File Type" on page 97

Hide quick links: