The NONAME00 increment-type file, starting with the 001 ID, is the active file by
default when you first use the 45MG, or after resetting the instrument memory.
Various types of files can be created, and IDs can be defined to represent various 1-D,
2-D, or 3-D thickness-measurement locations. When the instrument is restarted, it
automatically opens the last file used.
The following special cases may occur:
When the thickness value is blank, "———" is saved rather than a value.
When a measurement is already stored at the current ID, the new value
overwrites the old thickness reading unless the overwrite protection is enabled
(see "Setting the ID Overwrite Protection" on page 115).
When the ID increment reaches the end of a sequence and cannot be updated,
Last ID appears on the help text bar, a long beep sound is emitted (when the
beeper is active), and the ID on the display remains unchanged.
Refer to "Datalogger Option" on page 88 for more information on the datalogger.
To save data at the current ID in the active file
 While the desired thickness value and waveform are displayed, press [SAVE] to
save the measured thickness value.
OR
To save the measured thickness value and the waveform, set the SAVE KEY
DATA to THICKNESS + WF in the SYSTEM menu (see Figure 4-1 on page 44).