When DUAL XDCR TEST is selected, in the DUAL XDCR TEST screen (see
Figure 12-3 on page 205):
a) Press [MEAS] to start the test that measures the time of flight through each
side of a dual element transducer.
The measured parameter values appear.
b) Interpret the Tx and the Rx values as follows:
Similar values indicate a normal dual element transducer.
A difference between the values indicates that the delay line wear is not
the same for each element.
A missing value indicates that a cable is broken, or that the element is not
functioning.
c)
Take note of the calculated ZERO VALUE.
d) Press [ENTER] to terminate the dual element transducer test.

12.6 Performing the Software Diagnostic Test

The software diagnostic (SW DIAG) function generates an error log documenting all
errors that occurred during instrument operation. Olympus uses this information to
troubleshoot the 45MG operating software.
To access the software diagnostic
1.
In the measurement screen, press [SETUP], and then select SP MENU.

Figure 12-3 The DUAL XDCR TEST screen

Maintaining and Troubleshooting the 45MG
DMTA-10022-01EN, Rev. C, January 2015
205