Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015
2.
Press
to start the instrument.
The measurement screen appears. With a standard D79X dual element
transducer, the "Do--" message appears in the measurement screen (see
Figure 5-2 on page 52).

Figure 5-2 Initial screen with a standard D79X dual element transducer

3.
For dual element transducers, perform the transducer zero compensation:
a) Wipe the couplant off the tip of the transducer.
b) Press [2nd F], [CAL ZERO] (Do ZERO).
52
Chapter 5

Figure 5-1 Plugging in the transducer

Standard dual element
transducer connector
T/R 1 connector for a