The allowable number of characters for each custom point depends on the number of
ID characters defined in the start and end ID values. The total number of characters of
the ID value plus the custom points cannot exceed 20. For example, when the start
and end ID values are seven characters long, as shown in the example in Table 9 on
page 97, the maximum allowable length for each custom point is thirteen (20 − 7 = 13).
To create a sequential data file with custom points
1.
In the measurement screen, press [FILE], and then select CREATE (see "Creating
a Data File" on page 92 for details on the first parameters).
2.
At the bottom of the CREATE screen, select CONTINUE.
3.
In the second page of the CREATE screen (see Figure 6-20 on page 98):
a) Enter the START ID and END ID values.
b) Enter two or more CUSTOM POINTS values.
c)
Press [2nd F], [] to finish entering CUSTOM POINTS values.
d) Select CREATE.

Figure 6-20 Configuring ID range for a sequential with custom points data file type

6.7.2.5

2-D Grid Data File Type

A 2-D grid is a sequence of ID numbers arranged to describe a path in two
dimensions. Each part of the ID number corresponds to a particular matrix
dimension.