Figure 5-11 Coupling A Dual Element Transducer; Dmta-10022-01En, Rev. C, January - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015

NOTE
In general, use a thinner couplant (such as propylene glycol, glycerin, or water) for
smooth material surfaces. Rough surfaces require a more viscous couplant, such as
gel or grease. Special couplants are required for high-temperature applications.
2.
Using moderate to firm pressure, couple the tip of the transducer to the surface of
the test material, and keep the transducer as flat as possible on the material
surface (see Figure 5-11 on page 64).
3.
Read the measured thickness value for the tested part (see Figure 5-12 on
page 65).
64
Chapter 5

Figure 5-11 Coupling a dual element transducer

12.7 mm (0.500 in.)