Figure 7-7 Setting Up The Standard Alarm - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

2.
In the ALARM screen (see Figure 7-7 on page 135):
a) Set ALARM ENABLE to ON to activate the alarm function.
b) In ALARM MODE, select the desired alarm mode (STANDARD, B-SCAN,
or REDUCTION RT [reduction rate]).
The other parameters vary depending on the alarm mode selection.
NOTE
The B-SCAN option only appears when the B-scan mode is active (see "Using the
B-Scan" on page 156).
The REDUCTION RT option only appears when the active file is configured with the
FILE DATA MODE parameter set to REDUCTION RT (see "File Data Modes" on
page 104).

Figure 7-7 Setting up the STANDARD alarm

3.
When ALARM MODE is set to STANDARD or B-SCAN, set the LOW ALARM
and the HIGH ALARM values.
OR
When ALARM MODE is set to REDUCTION RT, set the YELLOW ALARM and
the RED ALARM values.
4.
Press [MEAS] to return to the measurement screen.
DMTA-10022-01EN, Rev. C, January 2015
Using Special Functions
135