Chapter 4: Installation
Additional Installation Equipment
To install the InnovOx On-Line TOC Analyzer, you will also need to provide the following equipment:
•
Mounting hardware to support the Analyzer (See
page
73.)
•
Mounting hardware to bolt Mounting Stand to wall or floor (if you ordered a Mounting Stand accessory)
•
Insulated wire (22-12 American Wire Gauge - AWG, rated to 300 Volts) for analog and alarm outputs
(optional; see
"Step 6: Install Input/Output, Cables, and Accessories" on page
•
Analog (4-20 mA) recorder (optional)
•
Insulated wire (12-12 American Wire Gauge, rated to 300 Volts) for AC power conduit (see
on Power and Configure Basic Settings" on page
•
An external circuit breaker or switch that disconnects both poles of the supply voltage, rated
appropriately
Step 2: Complete the Identification Records
Complete the Identification Records section on
installation and the Analyzer serial number (found on the side of the Analyzer).
Step 3: Select a Location for the Analyzer
Select a location in a well-ventilated area to avoid inhaling any toxic gases that may be emitted from the
Analyzer's waste. Place the Analyzer in a location away from direct sunlight and exposure to extreme
temperatures.
Operating at elevated temperatures (greater than 40 °C) prevents proper operation, and operating at low
temperatures (below 10 °C) can cause errors in the measurements or damage to the Analyzer.
Mounting Stand Installation
The InnovOx On-Line TOC Analyzer is designed to be mounted on a Mounting Stand or on a wall. For an
illustration of the required clearances for the Mounting Stand accessory,
GE Analytical Instruments ©2016
"Step 3: Select a Location for the Analyzer" on
102.)
"Identification Records" on page 3
WARNING
73 of 270
79.)
by recording the date of
see Figure
2.
DLM 68100-09 EN Rev. A
"Step 11: Turn