Calibrating The Analyzer; Table 9 Approximate Number Of Calibration Repetitions Possible - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 5: Calibration and Verification
7.
Enter a name for the new calibration protocol, and then press the Save button.
These ranges are based on a completely filled 40-mL Vial and no sample flushes. The maximum
number of repetitions that can be selected is 20. Divide the number of reps shown above by 2 for TOC
and TC modes.
Table 9: Approximate Number of Calibration Repetitions Possible
Mode
100 ppm
NPOC
TOC
IC
TC

Calibrating the Analyzer

To calibrate the Analyzer
1.
Select the Cal/Ver tab.
2.
Press the Calibrate button.
3.
Select a calibration protocol from the list by touching the screen. If you want to create a new
calibration, follow the previous steps in
4.
Press the Start Cal button and follow the on-screen prompts. The prompts will vary depending on the
settings you have specified in the selected calibration.
5.
After the Analyzer has completed measurements on all standards vials, the Calibration Summary screen
displays the following values. (Refer to
• Expected - Indicates the expected concentration of the calibration point.
When Auto Dilution is Off, the expected concentration is the Certified Concentration of the standard.
For reagent water, there is no expected concentration, so the entry is "N/A."
When Auto Dilution is On, the Expected concentration is the concentration of the standard after
dilution. For example, if a 1,000 ppm TOC standard is used in the calibration, and is diluted to
calibrate the Analyzer at 100 ppm, the Expected concentration will be 100 ppm.
• Measured: Indicates the average measured concentration of the standard, using default calibration
constants.
• RSD: Indicates the percent relative standard deviation of the non-rejected measurements of the
standard.
• Adjusted: Indicates the calculated concentration of the standard, using the new calibration
constants from the calibration just completed.
GE Analytical Instruments ©2016
1,000 ppm
6
7
3
3
6
7
3
3
"To set up a new calibration" on page
Figure 36 on page
Range
5,000 ppm
20,000 ppm
14
12
14
12
120.)
118 of 270
50,000 ppm
20
20
20
20
20
20
20
20
116.
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents