Chapter 4: Installation

Installing the Alarms and System Fault Output Connection

Four user-configured alarms are provided on TB3 and TB4. They may be assigned to any of the sample streams.
The concentration at which the alarm is triggered is set as described in
page
160.
The fifth alarm (the "System Fault Alarm") is configured to transmit an alarm whenever an Error is detected in
the Analyzer.
All of the alarms can be wired in both Normally Open (NO) and Normally Closed (NC) configurations.

Installing the Ethernet Connection

The Ethernet connection provides an additional means of capturing data from the Analyzer. Using Modbus,
commands may be issued to the Analyzer as well as collecting data. Refer to
and Modbus" on page 160
The Ethernet cable may be connected directly to a computer if a cross-over cable is used
the Analyzer to a
Computer). Or, as shown in
to a network. Cables are provided with the Analyzer.
GE Analytical Instruments ©2016
for details on the use of Modbus.
see Figure
Figure 10: Connecting the Analyzer to a Computer
"Configuring Alarm Outputs" on
11, a straight-through Ethernet cable is used to connect
86 of 270
"Using the Ethernet Connection
(Figure 10: Connecting
DLM 68100-09 EN Rev. A