Figure 33 Two-Point Calibration (Point 1 Is Reagent Water) - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 5: Calibration and Verification
The calibration of the Analyzer is defined by the intercept, a, and the slope, b.
For the case in which point 1 is reagent water, the calibration of the Analyzer is depicted graphically in
The operator designates point 1 as reagent water in the protocol by setting its concentration to 0 ppm. The
Analyzer interprets "0 ppm" as "reagent water."
When the Analyzer measures the reagent water, it also measures the TOC in the reagents and dilution water
used in the measurement. Because the TOC concentration in the reagent water is very low, this measurement
approximates the TOC concentration in the reagents and dilution water for the particular volumes of reagents
and dilution water used in the calibration. The Analyzer corrects the calibration for the TOC in the reagents and
dilution water by setting the measurement measured response (i.e., Mass Response) to equal 0 ppm.
The operator enters the concentration of point 2 into the calibration protocol. The concentration of the standard
used for point 2 is referred to as C
100 ppm, so C
= 100 ppm. C
2
The calibration constant, b, is calculated from the difference in the response to the two points:
where,
R
= Response of Analyzer to point 2
2
R
= Response of Analyzer to reagent water
RW
C
= TOC concentration of point 1 (= 0 for RW)
1
C
= TOC concentration of point 2 (mg/L)
2
GE Analytical Instruments ©2016
Figure 33: Two-Point Calibration (Point 1 is Reagent Water)
. In the example in Figure 16, the operator set the concentration of point 2 to
2
(or C
) for reagent water is zero.
1
RW
b =
110 of 270
C
- C
2
RW
R
- R
2
RW
Figure
33.
DLM 68100-09 EN Rev. A