Setting Up Input And Output (I/O); The I/O Tab; Connecting The Analyzer To A Network Via Ethernet; Configuring The Analyzer's Ip Address - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation

Setting Up Input and Output (I/O)

The I/O Tab

Connecting the Analyzer to a Network Via Ethernet

The Analyzer can be connected to your network or directly to a personal computer via an Ethernet cable. If you
connect the Analyzer to your network, use a standard Ethernet cable. If you connect the Analyzer directly to a
personal computer, use an Ethernet crossover cable.
N
:
Consult with your Information Technology group to determine the best configuration for your
OTE
networking environment.

Configuring the Analyzer's IP Address

You must configure the Analyzer's Internet Protocol (IP) settings for use in connecting the Analyzer to a
computer network or directly to a computer.
To configure the Analyzer's IP Address
1.
Select the I/O tab, and then press the IP Configuration button. The Configuration screen appears.
GE Analytical Instruments ©2016
Figure 61: IP Configuration Screen
156 of 270
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents