Collecting Historical And Real-Time Data; To Collect Real-Time Data (W/Ethernet Connection) - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation
11. Click the Close button.
12. Shut down, power off, and restart the computer.
To manually set the IP address for Windows
1.
On the computer, select Start Menu
2.
Next to the listing for the network connection, click View Status.
3.
Click the Properties button.
4.
If you are prompted to continue, click the Continue button.
5.
Scroll through the list and select Internet Protocol (TCP/IP)
6.
Click the Properties button.
7.
Select the radio button for "Use the following IP address"
8.
Enter an IP address. The first three numbers must be the same as those used on the Analyzer, and the
last number must be different from the one used on the Analyzer. Do not use the number 0 or 255.
For example, if the Analyzer IP address is set to the default of 192.168.1.10, you could set the computer
to 192.168.1.11.
9.
Enter a subnet mask address. This should be the same numbers as those used on the Analyzer.
10. Click the OK button.
11. Click the OK button.
12. Click the Close button.
13. Shut down, power off, and restart the computer.

Collecting Historical and Real-Time Data

Collect historical or real-time measurement data from the Analyzer using Ethernet and a data acquisition
program. For additional information, see
data file collected into a computer per the instructions in
Spreadsheet."
When the real-time data is collected, it includes the information shown in

To collect real-time data (w/Ethernet connection)

Configure the Ethernet connection and then use a data acquisition program to capture a TCP/IP transmitted .csv
text file in real time. Examples include PuTTY, HyperTerminal, and Tera Term.
1.
For direct connection of the Analyzer to a PC, select the crossover Ethernet cable from the Accessories
Kit. Attach one end to the Ethernet port in the electronics area of the Analyzer, and the other end to an
Ethernet port on the PC.
GE Analytical Instruments ©2016
®
Vista


Settings
Control Panel
"Installing the Ethernet Connection" on page
"Appendix C: Importing Analyzer Data into a
159 of 270
Network and Sharing Center.
86. Next, import the .csv
Table 13 on page