GE Sievers Innovox Operation And Maintenance Manual page 209

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 9: Troubleshooting
damage. 4) Thoroughly dry the sensor, the mount, and surrounding surfaces. 5) Reinsert the sensor into its
mount. 6) Using the Analyzer's touch panel, reset the error (Maintenance tab > Errors/Warnings button).
6443 - Fluid Leak in Optional C1D2 or ATEX Enclosure Error
Message: "Fluid leak detected in optional outer enclosure. Resolve the source of the leak. Clean and dry the
sensor.
Explanation: The liquid sensor is a corrosion-resistant and optically-based sensor intended for detecting large
leaks inside the optional outer enclosure. The sensor is mounted on the lower-left inside wall of the outer
enclosure and triggers when the face of the sensor contacts fluid. (Due to variations in installation conditions,
fluid from leaks may not immediately contact the sensor.) When a leak is detected, the Analyzer reports this
error, displays a pop-up message, and stops analysis. Do the following: 1) Identify and repair the source of the
leak. 2) Wipe up the fluid inside the outer enclosure. 3) Remove the sensor from its mount, making sure to pull it
straight up to release and protect it from damage. 4) Thoroughly dry the sensor, the mount, and surrounding
surfaces. 5) Reinsert the sensor into its mount. 6) Using the Analyzer's touch panel, reset the error (Maintenance
tab > Errors/Warnings button).
GE Analytical Instruments ©2016
209 of 270
DLM 68100-09 EN Rev. A