Installing The Analysis Complete Output Connection; Figure 8 Wiring Option - Binary Input Using Analyzer's Internal Supply; Figure 9 Wiring Option - Binary Input Using Analyzer's External Supply - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 4: Installation
Figure 8: Wiring Option — Binary Input using Analyzer's Internal Supply
Figure 9: Wiring Option — Binary Input using Analyzer's External Supply

Installing the Analysis Complete Output Connection

The analysis complete (binary output) on TB3 may be used as a control signal to a sample sequencer device.
Once the binary output is wired properly, the Analyzer will automatically provide a one-second contact closure
at the end of every sample analysis.
No menu configuration is necessary. The binary output is active in the Online, Check Standard, and Grab modes.
GE Analytical Instruments ©2016
85 of 270
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents