Figure 108 Binary Input Check - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 9: Troubleshooting
If a leak sensor is installed in the Analyzer and the leak sensor is enabled, the Main Enclosure Leak Sensor
button appears with either a No Liquid Detected or Liquid Detected state. To test an auxiliary leak sensor
installed in an ATEX enclosure, first access the Liquid Sensor Setup screen (I/O tab
Liquid Sensor Setup) and
temporarily disable the Main Enclosure Leak Sensor button. On the Binary Input screen (Maintenance tab
Diagnostics), press the Binary Input button. The Auxiliary Enclosure Leak Sensor button appears with either
a No Liquid Detected or Liquid Detected state. Before continuing with Diagnostics, re-enable the Main Enclosure
Leak Sensor button (I/O tab
Liquid Sensor Setup).
Figure 108: Binary Input Check
GE Analytical Instruments ©2016
241 of 270
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents