Chapter 9: Troubleshooting
If a leak sensor is installed in the Analyzer and the leak sensor is enabled, the Main Enclosure Leak Sensor
button appears with either a No Liquid Detected or Liquid Detected state. To test an auxiliary leak sensor
installed in an ATEX enclosure, first access the Liquid Sensor Setup screen (I/O tab
Liquid Sensor Setup) and
temporarily disable the Main Enclosure Leak Sensor button. On the Binary Input screen (Maintenance tab
Diagnostics), press the Binary Input button. The Auxiliary Enclosure Leak Sensor button appears with either
a No Liquid Detected or Liquid Detected state. Before continuing with Diagnostics, re-enable the Main Enclosure
Leak Sensor button (I/O tab
Liquid Sensor Setup).
Figure 108: Binary Input Check
GE Analytical Instruments ©2016