Analog Outputs (Tb1); Stream Id - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 4: Installation
N
:
4-20 mA outputs are isolated from system ground.
OTE
4-20 mA outputs are not individually isolated from each other and, therefore, are not intended to be
connected to multiple systems with differences in ground potential.
4-20 mA outputs may be configured by the user to output NPOC, IC, TC, or TOC ppm levels.
Active Stream
Standby
Error
Warning
GE Analytical Instruments ©2016
Table 3: Analog Outputs (TB1)
Pin
Signal
1
Analog 1 (+)
2
GND_ISO (-)
3
ANALOG 2 (+)
4
GND_ISO (-)
5
ANALOG 3 (+)
6
GND_ISO (-)
7
ANALOG 4 (+)
8
GND_ISO (-)
9
ANALOG 5 (+)
10
GND_ISO (-)
11
ANALOG 6 (+)
12
GND_ISO (-)
Table 4: Stream ID
Output matches the standby level set for the first
sequentially-ordered active stream, starting with Stream 1
1
2
3
4
5
Output matches the error level set for the first
sequentially-ordered active stream, starting with Stream 1
n/a (Does not change current value.)
Function
4-20mA Stream #1
4-20mA Stream #2
4-20mA Stream #3
4-20mA Stream #4
4-20mA Stream #5

Stream ID

(11 mA, 12 mA, 13 mA, 14 mA, or 15 mA, as
designated in
Table 4: Stream ID
Stream ID
11 mA
12 mA
13 mA
14 mA
15 mA
81 of 270
DLM 68100-09 EN Rev. A