Chapter 6: Analyzer Operation
3.
Touch the protocol name on the screen.
4.
To modify this protocol, press the Modify button. The Grab (Setup) screen appears.
5.
Make changes to the protocol, as needed:
• Press the Measure button to specify which value to measure: IC, NPOC, TOC, or TC.
• Press the Range button to specify one of the five range buttons, or select the Auto button for the
Auto Range function (see
• Press the Acid and Oxidizer buttons to specify the percent of each reagent to use in the protocol.
• Press the Reps button to change the default setting for the number of measurements (repetitions)
the Analyzer will make for this sample. If Repeat Criteria is ON, then this field will be grayed out and
uneditable.
• Press the Rejects button to change the default setting for the number of measurements the
Analyzer should exclude from its calculations. If Repeat Criteria is ON, then this field will be grayed
out and uneditable.
N
:
number of rejects must be less than the number of repetitions.
The
OTE
• Press Repeat Criteria to allow the Analyzer to follow configurable logic to determine the number of
repetitions to run. If ON is selected, you will be able set the maximum number and minimum number
of repetitions for the vial. The number of reps will be determined by the InnovOx depending on what
you enter for the RSD criteria. If you enter 7 Max Reps and 2 Min Reps, the Analyzer will run at least
two repetitions and compare the calculated RSD to the RSD criteria entered. If the calculated RSD is
GE Analytical Instruments ©2016
Figure 51: The Grab (Setup) Screen
"Auto Range" on page
144).