Unscrewing The Syringe Base; Figure 75 Locating The Syringe Screw And Washers - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 8: Maintenance
5.
Using a 3/32 Allen wrench, loosen and remove the screw that secures the syringe. Remove any washers
under the block, and retain them for re-installation
6.
Unscrew the syringe base by turning the base to the left.
GE Analytical Instruments ©2016
Figure 75: Locating the Syringe Screw and Washers
Figure 76: Unscrewing the Syringe Base
185 of 270
Washers*
*If these washers are NOT installed
with your Analyzer. ignore this step.
DLM 68100-09 EN Rev. A