Chapter 9: Troubleshooting
6417 – Stream 2 Flow Warning
Message: "Stream 2 flow stopped."
Explanation: Check that water is flowing in the sample cup and the inlet valve (if installed) is properly adjusted. If
the problem persists, please export the warnings and contact GE Analytical Instruments for further instructions.
6418 – Stream 3 Flow Warning
Message: "Stream 3 flow stopped."
Explanation: Check that water is flowing in the sample cup and the inlet valve (if installed) is properly adjusted. If
the problem persists, please export the warnings and contact GE Analytical Instruments for further instructions.
6419 – Stream 4 Flow Warning
Message: "Stream 4 flow stopped."
Explanation: Check that water is flowing in the sample cup and the inlet valve (if installed) is properly adjusted. If
the problem persists, please export the warnings and contact GE Analytical Instruments for further instructions.
6420 – Stream 5 Flow Warning
Message: "Stream 5 flow stopped."
Explanation: Check that water is flowing in the sample cup and the inlet valve (if installed) is properly adjusted. If
the problem persists, please export the warnings and contact GE Analytical Instruments for further instructions.
6421 – Over Range Error
Message: "The measured value was too high for the current range."
Explanation: Select an appropriate higher range and restart the analysis.
6422 – Gas Inlet Pressure Error
Message: "Gas inlet pressure is low."
Explanation: Check source gas. Be sure that your lines are connected and that there is sufficient pressure. If
using the air purifier, check to be sure it is powered and functioning. If the problem persists, please export the
errors and contact GE Analytical Instruments for further instructions.