Chapter 2: System Description
Material
®
AFLAS
Norprene Chemical
with PVC lining
®
PEEK
PVC
PCTFE, PTFE
®
Kalrez
Additional System Components
System/Controller and Electronics
Operation of the Analyzer is controlled by an advanced X86 microprocessor on the System/Controller Board,
which communicates with the other Analyzer modules over a Multi-point Low Voltage Differential Signaling
(M-LVDS) bus. Other electronic assemblies provide the interface, monitor and control functions for the Analyzer:
• Color LCD VGA display with touch panel
• Multistream module board for microprocessor-controlled selection of the sample to be measured
(This board also controls operation of the optional sample pump accessory.)
• Sample handling module board for microprocessor-controlled preparation of the sample,
reagents, and dilution water
• Gas control module board for microprocessor-controlled control of the carrier and sparge gases
• Reactor module board for microprocessor-controlled oxidation of the samples
• NDIR module board for microprocessor-controlled measurement of the CO
Sample Handling and Reactor Modules
• Input/Output board for processing digital input signals and analog and digital outputs
• Terminator board, which terminates the M-LVDS bus and provides Ethernet and USB interfaces
• Power Entry/Supply
GE Analytical Instruments ©2016
Table 1: InnovOx Wetted Materials List
O‐rings on seals
Waste tubing
Sample manifold, reagent inlet fittings
Gas‐liquid separator, NDIR detector, sparger,
multistream manifold, sample cups, external
tubing
Vial port valve
Reagent valves
and PEEK
Components
60 of 270
produced in the
2
DLM 68100-09 EN Rev. A