Figure 62 Alarm Output Screen - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation
3.
Press the Alarm Stream button. Buttons for each stream appear to the right.
4.
Press the stream button you want to the alarm to respond to (such as the Stream 1 button or Stream 2
button).
5.
Press the High Level Alarm button and use the numeric keypad that appears to the right to set the
threshold concentration (in ppm units
6.
Press the Low Level Alarm button and use the numeric keypad that appears to the right to set the
threshold concentration (in ppm units
N
:
The low-level set point value must be less than the high-level set point value.
OTE
7.
To activate the alarms, press the Alarm On/Off button, and then press the On button that appears.
N
:
The high and low threshold concentrations refer to the parameter the stream is configured to
OTE
measure. For example, if Alarm 1 responds to Stream 1, and Stream 1 is configured to measure
TOC, then the High Level Alarm and the
10.
The unit of measure appears in the default ppm unit or another unit measure if the stream's Setup screen
has been configured with a conversion factor.
GE Analytical Instruments ©2016
Figure 62: Alarm Output Screen
10
). Any measurement above this threshold will trigger the alarm.
10.
). Any measurement below this threshold will trigger the alarm.
Low Level Alarm
161 of 270
are in the units of ppm TOC.
DLM 68100-09 EN Rev. A